FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3873795 · Received January 27, 2014

Report

Report Number
1314492-2014-04665
Event Type
Malfunction
Date Received
January 27, 2014
Date of Event
December 1, 2013
Report Date
December 31, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURER REF. NO.: (B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE REPORTED SYMPTOM "DOOR NOT FULLY LATCHED" CAUSED BY A BROKEN DOOR HOOK. THE DOOR HOOK WAS REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED "DOOR NOT FULLY LATCHED". IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
59346 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1