2.15MM X 22MM SPIRAL ROUTER
Report
- Report Number
- 1045834-2013-16430
- Event Type
- Malfunction
- Date Received
- November 27, 2013
- Report Date
- June 22, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE WAS RECEIVED AND EVALUATED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE CUSTOMER'S COMPLAINT OF PACKAGING DEFECTS COULD NOT BE CONFIRMED. VISUAL INSPECTION CRITERIA ARE "NO LOOSE FOREIGN MATERIAL (LFM) WHEN INSPECTED AT 1X MAGNIFICATION". NO LFM COULD BE FOUND ON DEVICE WHEN INSPECTED AT 10X MAGNIFICATION. ALSO, THE PACKAGE IS IN GOOD CONDITION WITH NO DEFECTS OF THE PEELABLE OR TERMINAL SEAL. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
REPORT 4 OF 6. REPORT RECEIVED FROM JAPAN STATING THAT THE DEVICE WAS RETURNED DUE TO ¿RETURNED PRODUCT UNUSED ¿ INCOMING ORDER FAILED DISTRIBUTOR'S INSPECTIONS¿. THE DEVICE WAS NOT BEING USED DURING SURGERY AND NO INJURY OR MEDICAL INTERVENTION OCCURRED. THE DATE OF EVENT IS UNKNOWN. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 619617 | 2.15MM X 22MM SPIRAL ROUTER | HBC | HBC | DEPUY SYNTHES POWER TOOLS | E473057911 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |