SPECTRANETICS LEAD LOCKING DEVICE
Report
- Report Number
- 1721279-2014-00096
- Event Type
- Malfunction
- Date Received
- June 14, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 4, 2014
- Manufacturer
- SPECTRANETICS CORPORTATION
- Product Code
- DRB
- PMA / PMN Number
- K043401
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
REPORT FOR THE GLIDELIGHT IS MDR REPORT 1721279-2014-00095.
LEAD MANAGEMENT CASE TO EXTRACT TWO CARDIAC LEADS DURING A GEN CHANGE. THE FIRST LEAD WAS A 1581 RIATA (IMPL. (B)(6) 2006 IN THE RV), DUE TO EXPOSED CABLES THE DECISION WAS MADE TO USE A 16F GLIDELIGHT FOR EXTRACTION. AN LLD EZ WAS USED TO PREP THE LEAD AND LASING BEGAN WITH THE GLIDELIGHT. THE GLIDELIGHT WAS MOVING SLOWLY AND SMOOTHLY WITH NO INDICATION OF A COMPLICATION, UPON ENTERING THE RA PROGRESS WAS HALTED AND IT WAS NOTICED UNDER FLUOROSCOPY THAT THE LEAD STARTED TO BUNCH UP. A LITTLE MORE PROGRESS WAS ATTEMPTED BEFORE DECIDING TO CHANGE OVER TO THE RA LEAD (SJM VENTRATEK MODEL NUMBER UNKNOWN IMPLANTED (B)(6) 2006) WHEN THE BLOOD PRESSURE DROPPED. THE SURGEON IMMEDIATELY PERFORMED A STERNOTOMY AND ON EXAMINATION OF THE SITE IT WAS DISCOVERED THAT THE ENTIRE INNOMINATE WAS TORN. IT WAS BELIEVED BY THE PHYSICIAN THAT THE LEAD HAD BEEN EXTRA-CARDIAC PROVIDING AN EXTERNALIZED TRACK FOR THE GLIDELIGHT TO FOLLOW CAUSING THE TEAR. THE INNOMINATE INJURY WAS REPAIRED AND EXTRACTION ATTEMPTS WERE STOPPED FOR FEAR OF CAUSING THE VESSEL TO TEAR FURTHER. THE LLD EZ WAS CUT AND CAPPED WITHIN THE LEAD. THE PATIENT SURVIVED INTERVENTION. THIS REPORT IS TO REFLECT THE ABANDONMENT OF THE LLD EZ.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351852 | SPECTRANETICS LEAD LOCKING DEVICE | LLD EZ | DRB | SPECTRANETICS CORPORTATION | 518-062 | FLP14A14A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other | SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM| SPECTRANETICS 16F GLIDELIGHT LASER SHEATH| SJM 1581 RIATA CARDIAC LEAD (IMPL. (B)(6) 2006)| SJM VENTRATEK (UNK MODEL # IMPL. (B)(6) 2006) |