FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3873424 · Received June 13, 2014

Report

Report Number
2531779-2014-17050
Event Type
Malfunction
Date Received
June 13, 2014
Report Date
June 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 07/22/2014, DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH NO VISIBLE DAMAGE TO THE KEYPAD. DURING TESTING, THE UP ARROW AND DOWN ARROW BUTTONS WERE INTERMITTENTLY RESPONSIVE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL OF THE KEYPAD BUTTON CONTACTS. UNRELATED TO THE INITIAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED ALONG THE THREADS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350214 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 61 YR