FDA Adverse Event Malfunction Summary report: N

PRECISION®

MDR report key: 3873388 · Received June 13, 2014

Report

Report Number
3006630150-2014-01346
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 21, 2014
Report Date
May 22, 2014
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
030017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS CHARGING THE IPG FREQUENTLY. THE PATIENT HAD A PREVIOUS REVISION WHEREIN ELECTROCAUTERY WAS USED. MONOPOLAR ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND CURRENT COMPANY LABELING WARNS AGAINST THE USE OF MONOPOLAR ELECTROCAUTERY. (PHYSICIAN¿S IMPLANT MANUAL 9055940-001).

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THE PHYSICIAN WANTED TO UPGRADE THE PATIENT¿S IPG DUE TO THE AGE OF THE BATTERY. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND DID WELL POST OPERATIVELY. NO MALFUNCTION WAS SUSPECTED. THE EXPLANTED IPG WAS NOT RETURNED TO BSN AS IT WAS DISCARDED BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT¿S IPG WAS FAILING AND WAS NOT HOLDING A CHARGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT¿S IPG WAS FAILING AND WAS NOT HOLDING A CHARGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT¿S IPG WAS FAILING AND WAS NOT HOLDING A CHARGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350131 PRECISION® SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1110 NA