PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3873335
·
Received June 13, 2014
Report
- Report Number
- 3004209178-2014-85732
- Event Type
- Injury
- Date Received
- June 13, 2014
- Date of Event
- May 17, 2014
- Report Date
- May 27, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. UNIT WAS RECEIVED WITH A CRACKED DISPLAY WINDOW CORNER, SCRATCHED LCD WINDOW, AND A CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
THE CUSTOMER CALLED IN TO REPORT THAT SHE HAD BEEN HOSPITALIZED FOR HIGH BLOOD GLUCOSE LEVELS. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 950 MG/DL. THE CALLER ALSO STATED THAT SHE RECEIVED A MOTOR ERROR ALARM. NO FURTHER INFORMATION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349563 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Hospitalization |