FDA Adverse Event Malfunction Summary report: N

VERSALET INFANT INCUBATOR/RADI P84

MDR report key: 387327 · Received April 6, 2002

Report

Report Number
2510954-2002-00002
Event Type
Malfunction
Date Received
April 6, 2002
Date of Event
March 6, 2002
Report Date
March 18, 2002
Manufacturer
HILL-ROM AIRSHIELDS
Product Code
FMZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

2002 SITUATION WAS DESCRIBED: THE BED HAS BEEN TOO WARM IN INCUBATOR MODE, NURSE HAD TURNED UNIT OFF TEMPORARILY. RESP THERAPIST AND X-RAY TECH WANTED TO MAKE MORNING X-RAY. PLATFORM ASSEMBLY AND OHA WERE IN THE LOWEST POSITION (INCUBATOR). THEY DECIDED THEY NEEDED TO RAISE OHA TO COMPLETE X-RAY, PUSHED UP ARROW ON OHA, UNIT DID NOT RAISE, THEN NOTICED POWER WAS OFF, THEN POWERED UNIT ON. UNIT FINISHED SELF CHECK THEN SPONTANEOUSLY, WHOLE PLATFORM WENT UP, NURSE PUSHED DOWN ARROW TO STOP UNIT FROM RAISING, UNIT WOULD NOT STOP. AS UNIT WENT UP THE SHELF CAUGHT ON THE EPICARE POWER COLUMN THEN BROKE OFF. THE SHELF WAS HOLDING A SATURATION TCM, A TRANSCUTARIOUS MONITOR AND CALIBRATION TANKS (MEDGAS). TECH CAUGHT THE SHELF AS IT BROKE OFF PREVENTING CONTENTS FROM FALLING. VENTILATOR TUBING WAS REACHING FULL EXTENSION CAPABILITY WHEN CHARGE NURSE UNPLUGGED UNIT AND IT STOPPED RAISING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSALET INFANT INCUBATOR/RADI P84 COMBINATION INCUBATOR/WARMER FMZ HILL-ROM AIRSHIELDS 7700 CARE CENTER *

Patients

Seq Age Sex Outcome Treatment
1 *