FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3873105 · Received June 13, 2014

Report

Report Number
2955842-2014-03649
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
April 24, 2014
Report Date
May 15, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS WAS UNABLE TO REPLICATE THE CUSTOMER REPORTED COMPLAINT. FAILURE ANALYSIS FOUND THAT THE INSTRUMENT PASSED A CUT TEST ON AN IN-HOUSE IS3000 SYSTEM. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE FOLLOWING ADDITIONAL DAMAGES TO THE INSTRUMENT: THE INSTRUMENT MAIN TUBE WAS CRACKED. THE TUBE HAD A SMALL FISSURE IN THE AXIAL DIRECTION, LOCATED DIRECTLY BELOW THE TUBE REINFORCEMENT RING. THE LOCATION OF THE FISSURE WAS IN AN AREA THAT IS NOT PROTECTED BY TIP COVER. THE LOCATION AND TYPE OF MAIN TUBE CRACKING IDENTIFIED IN THIS REPORT DIFFERS FROM MAIN TUBE CRACKS CAPABLE OF LEAKING ELECTROSURGICAL ENERGY. THE LOCATION OF THESE CRACKS ON THE INSTRUMENT MAIN TUBE IN THIS REPORT IS LIMITED TO AN AREA WHERE THE MAIN TUBE OF THE INSTRUMENT COVERS THE EXTENSION INSERT, ESSENTIALLY PROVIDING DOUBLE INSULATION OF THE TUBE, WITH ADDITIONAL SILICONE SEAL ADHESIVE BONDING. THERE IS NO PARTING OF MATERIAL IN THESE CASES AND THERE IS NO PROPAGATION OR GROWTH OF THE CRACKS. THE CRACKS APPEAR TO BE A RESULT OF LEVERAGE FORCE CAUSED BY EXPOSING THE INSTRUMENT TIP TO EXCESSIVE SIDE LOAD FORCE OR MISHANDLING (FOR EXAMPLE INSTRUMENT DROPS). THERE ARE NO REPORTED CASES OF ENERGY LEAKAGE FROM CRACKS IN THIS AREA. PAD PRINTING WAS REMOVED FROM THE DISTAL END OF THE TUBE EXTENSION. THE REMOVED AREA MEASURED APPROXIMATELY .396 X .237. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE PAD PRINTING REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS DULL DURING A DA VINCI PROSTATECTOMY PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349480 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-12 M10140123 214

Patients

Seq Age Sex Outcome Treatment
1