PENUMBRA COIL 400
Report
- Report Number
- 3005168196-2014-00381
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- PENUMBRA, INC.
- Product Code
- HCG
- PMA / PMN Number
- K120330
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NURSE
Narratives
CONCLUSION: THE COMPLIANT HAS BEEN EVALUATED. THE COMPLAINT INDICATES THAT THE COIL WAS CATCHING ON THE EDGE OF THE PXSLIM CATHETER WHILE ADVANCING THE COIL OUT OF THE DISTAL TIP. IT ALSO STATES THAT THE COIL WAS ADVANCED AND RETRIEVED ON THE STERILE BENCH OUTSIDE THE PATIENT AND CATCHING CONTINUED. EVALUATION OF THE RETURNED DEVICE REVEALED A KINK IN THE PROXIMAL END OF THE PUSHER ASSEMBLY, LIKELY RELATED TO REPEATED ATTEMPTS TO PUSH THE COIL THROUGH THE CATHETER. IN ADDITION, THE DISTAL TIP OF THE PXSLIM MICROCATHETER IS OVALIZED. THE LIKELY CAUSE OF THE ORIGINAL COMPLAINT IS DEFORMATION (OVALIZATION) OF THE DISTAL TIP OF THE MICROCATHETER CAUSING THE LUMEN OF THE CATHETER TO BE INCOMPATIBLE WITH THE OUTER DIAMETER OF THE COIL WHICH FELT AS THOUGH IT WAS "CATCHING" ON THE CATHETER AS DESCRIBED IN THE COMPLAINT. IT WAS NOTED DURING EVALUATION THAT THE CATHETER TIP WAS STEAM SHAPED BY THE USER. THE TIP MAY HAVE BEEN DEFORMED DURING THIS PROCESS. THESE DEVICES ARE 100% INSPECTED FOR DAMAGE DURING PACKAGING.
RESULT: THE COIL IS STILL ATTACHED TO THE PUSHER ASSEMBLY. THE PUSHER ASSEMBLY WIRE IS SLIGHTLY KINKED APPROXIMATELY 5.0 CM FROM THE PROXIMAL END OF THE WIRE. CONCLUSION: THE COMPLIANT HAS BEEN EVALUATED. THE COMPLAINT INDICATES THAT THE COIL WAS CATCHING ON THE EDGE OF THE PXSLIM CATHETER WHILE ADVANCING THE COIL OUT OF THE DISTAL TIP. IT ALSO STATES THAT THE COIL WAS ADVANCED AND RETRIEVED ON THE STERILE BENCH OUTSIDE THE PATIENT AND CATCHING CONTINUED. EVALUATION OF THE RETURNED DEVICE REVEALED A KINK IN THE PROXIMAL END OF THE PUSHER ASSEMBLY, LIKELY RELATED TO REPEATED ATTEMPTS TO PUSH THE COIL THROUGH THE CATHETER. IN ADDITION, THE DISTAL TIP OF THE PXSLIM MICROCATHETER IS OVALIZED. THE LIKELY CAUSE OF THE ORIGINAL COMPLAINT IS OVALIZATION OF THE DISTAL TIP OF THE MICROCATHETER CAUSING THE LUMEN OF THE CATHETER TO BE INCOMPATIBLE WITH THE OUTER DIAMETER OF THE COIL WHICH FELT AS THOUGH IT WAS "CATCHING" ON THE CATHETER AS DESCRIBED IN THE COMPLAINT. THESE DEVICES ARE 100% INSPECTED FOR DAMAGE DURING PACKAGING. THE OVALIZATION IN THE DISTAL TIP OF THE MICROCATHETER MAY HAVE BEEN RELATED TO HANDLING OF THE DEVICE DURING REMOVAL FROM THE PACKAGING AND PREPARATION FOR USE. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS. THIS MDR IS ASSOCIATED WITH MDR 3005168196-2014-00382.
THE PATIENT WAS UNDERGOING AN EMBOLIZATION PROCEDURE USING A PENUMBRA COIL 400 AND A PX SLIM DELIVERY MICROCATHETER. DURING THE PROCEDURE, THE PHYSICIAN FELT THE DISTAL END OF THE COIL 400 SNAG THE TIP OF THE PX SLIM MICROCATHETER. THIS CAUSED THE CATHETER TIP TO DISLODGE FROM THE COIL MASS. THE PX SLIM AND COIL 400 WERE REMOVED AND THERE WAS NO REPORT OF AN ADVERSE EFFECT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350099 | PENUMBRA COIL 400 | HCG, KRD | HCG | PENUMBRA, INC. | F38900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |