FDA Adverse Event Injury Summary report: N

RIATA ST OPTIM ACTIVE FIXATION

MDR report key: 3872986 · Received June 13, 2014

Report

Report Number
2938836-2014-11872
Event Type
Injury
Date Received
June 13, 2014
Date of Event
January 24, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WITH THE LEAD TIP MEASURING 34.5CM WAS RETURNED FOR ANALYSIS. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS EXPLANTED AND REPLACED DUE TO A RISE IN LEAD IMPEDANCES AND SUSPECTED LEAD FRACTURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350068 RIATA ST OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7020/65 NA

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention