DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 3872955
·
Received June 13, 2014
Report
- Report Number
- 2938836-2014-11833
- Event Type
- Injury
- Date Received
- June 13, 2014
- Date of Event
- April 26, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
A PARTIAL LEAD WAS RETURNED IN FIVE SEGMENTS FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL AFTER RECEIVING A PATIENT NOTIFICATION. UPON INTERROGATION HIGH, OUT OF RANGE, HIGH VOLTAGE LEAD IMPEDANCE WAS NOTED. PACING LEAD IMPEDANCE WAS NOTED TO HAVE DECREASED AND CAPTURE THRESHOLD SLIGHTLY INCREASED. LEAD FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. PATIENT WAS FINE AFTER THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349192 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7122/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |