FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 3872955 · Received June 13, 2014

Report

Report Number
2938836-2014-11833
Event Type
Injury
Date Received
June 13, 2014
Date of Event
April 26, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WAS RETURNED IN FIVE SEGMENTS FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL AFTER RECEIVING A PATIENT NOTIFICATION. UPON INTERROGATION HIGH, OUT OF RANGE, HIGH VOLTAGE LEAD IMPEDANCE WAS NOTED. PACING LEAD IMPEDANCE WAS NOTED TO HAVE DECREASED AND CAPTURE THRESHOLD SLIGHTLY INCREASED. LEAD FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. PATIENT WAS FINE AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349192 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7122/65 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention