FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3872860 · Received June 13, 2014

Report

Report Number
1416980-2014-18930
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
April 29, 2014
Report Date
May 21, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE DEVICE WAS RETURNED AND EVALUATED FOR THE REPORTED EVENT. A REVIEW OF THE EVENT HISTORY LOG VERIFIED THE OCCURRENCE OF AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. THE CAUSE WAS DETERMINED TO BE FALSE EMPTY DETECT AND USE ERROR, CLINICIAN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 09:15:48. DURING NIGHT DRAIN CYCLE SIX, THE PATIENT'S ULTRAFILTRATION READING WAS 1302ML, INDICATING THE HOME PATIENT (HP) DRAINED 1302ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1800ML. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350240 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1