FDA Adverse Event Malfunction Summary report: N

BLACKMAX-NEURO

MDR report key: 3872720 · Received June 13, 2014

Report

Report Number
1045834-2014-12264
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 1, 2014
Report Date
May 27, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE¿S POWER WAS BROKEN. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. IT WAS DETERMINED THAT THIS WAS DUE TO MISUSE BY RUNNING THE DEVICE IN THE SAFE OR LOAD POSITION. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE, ABUSE AND/OR USER ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MOTOR DEVICE ¿QUITS WHEN PRESSURE IS APPLIED TO THE BURR¿. THE EVENT WAS NOT REPORTED TO HAVE OCCURRED DURING SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO INJURIES, MEDICAL INTERVENTIONS OR PROLONGED HOSPITALIZATIONS REPORTED. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, THE REPORTER CLARIFIED THE EVENT OCCURRED IN (B)(6) 2014. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349598 BLACKMAX-NEURO MOTOR, DRILL, PNEUMATIC - HANDPIECE HBB DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 CUTTER DEVICE