FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3872660 · Received June 13, 2014

Report

Report Number
3005075853-2014-03987
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
June 6, 2014
Report Date
June 9, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE TISSUE PAD DAMAGED, MELTED, AND A SMALL PORTION WAS NOT RETURNED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT ACTIVATED DURING FUNCTIONAL TESTING. THERE WERE NO ALERT SCREENS AT ANY TIME DURING TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC RIGHT HEMICOLECTOMY PROCEDURE, THE DEVICE WAS NOT WORKING. SHOWN ERROR DURING THE SURGERY. TISSUE PAD CAME OUT WHILE OPERATING. NO PATIENT CONSEQUENCES REPORTED. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349281 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE