FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3872644 · Received May 23, 2014

Report

Report Number
3008642652-2014-01592
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
May 1, 2014
Report Date
May 22, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE BELT FAILED INCOMING FUNCTIONAL TESTING. INVESTIGATION REVEALED THERE WAS A SHORTED PULSE WIRE IN THE DISTRIBUTION NODE (DN) TO FRONT THERAPY ELECTRODE (TE) CABLE. THE ROOT CAUSE FOR THE SHORTED PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED PULSE WIRE. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. UPON SERVICING ELECTRODE BELT SN (B)(4), THE ELECTRODE BELT FAILED INCOMING FUNCTIONAL TESTING. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307579 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA