FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYS
MDR report key: 3872604
·
Received May 23, 2014
Report
- Report Number
- 3008642652-2014-01590
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- April 29, 2014
- Report Date
- May 22, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE BATTERY CHARGER/MODEM WAS RESETTING. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CONTAMINATION TO THE BATTERY BOARD CONTACTS. THE ROOT CAUSE OF THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY INGRESS OF AN UNK LIQUID. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/MODEM.
Description of Event or Problem · 1
A REVIEW OF THE SERVICE DATA DETECTED A REPORTABLE EVENT. UPON SERVICING BATTERY CHARGER SN (B)(4), THE CHARGER WAS RESETTING. THE LAST PT TO USE THIS CHARGER DID NOT REPORT ANY DEFICIENCIES.