FDA Adverse Event Malfunction Summary report: N

PLUM XLM W/DATAPORT

MDR report key: 3872490 · Received April 18, 2014

Report

Report Number
9615050-2014-02784
Event Type
Malfunction
Date Received
April 18, 2014
Date of Event
January 1, 2011
Report Date
September 14, 2011
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K010924
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

INITIAL TESTING AT THE USER FACILITY OCCURRED ON (B)(6) 2011. INITIALLY, THE AC POWER CORD WAS TO BE RETURNED TO THE SERVICE CENTER FOR FURTHER TESTING; HOWEVER, THE AC POWER CORD WAS NOT RECEIVED. HOSPIRA COULD NOT COMPLETE THE TESTING AND INVESTIGATION ON THE POWER CORD ASSOCIATED WITH THE REPORTED EVENT. THIS REPORT IS BEING MADE AS PART OF A RETROSPECTIVE REVIEW. NO FURTHER TESTING OR INVESTIGATION WILL BE PERFORMED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE GROUND PRONG ON THE AC POWER CORD WAS BROKEN. THE DEVICE HAD BEEN RETURNED TO THE BIOMEDICAL DEPT FOR A REPORT OF THE PUMP DOES NOT FUNCTION. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, DEVICE PROGRAMMING OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTED ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPY WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
238608 PLUM XLM W/DATAPORT 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK