FDA Adverse Event Malfunction Summary report: N

CONTOUR NEXT

MDR report key: 3872332 · Received June 13, 2014

Report

Report Number
1826988-2014-00202
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 1, 2014
Report Date
May 30, 2014
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K111268
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER RETURNED ONLY THE CONTOUR BLOOD GLUCOSE TEST STRIPS FOR EVALUATION. MODEL# 7080G, LOT# 3HJ3F02A, EXP DATE 08/31/2015; MANUFACTURE DATE 08/01/2013; 510(K) K062058.

Description of Event or Problem · 1

THE CUSTOMER RECEIVED BLOOD GLUCOSE READINGS OF 191,41 AND 93MG/DL ON THE CONTOUR NEXT EZ METER. HE ALSO RAN TESTS ON THE CONTOUR METER AND THE READINGS WERE 233, 50 AND 108MG/DL. THE DIFFERENCES BETWEEN THE READINGS FALL IN THE "C" AND "D" ZONES OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCES CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CONTOUR NEXT TEST STRIPS WERE EXPECTED TO BE RETURNED FOR EVALUATION BUT CUSTOMER RETURNED ONLY THE CONTOUR TEST STRIPS. REPLACEMENT TEST STRIPS AND METERS WERE SENT TO THE CUSTOMER.