FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 3872319 · Received June 13, 2014

Report

Report Number
2531779-2014-16983
Event Type
Malfunction
Date Received
June 13, 2014
Report Date
June 4, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 08/14/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE TEXT ON THE DISPLAY SCREEN WAS DIM/FADED AND DISCOLORED. UNRELATED TO THE DISPLAY SCREEN, THE PUMP CASE WAS DAMAGED WHERE THE LOW PROFILE CLIP ATTACHED TO THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, A REPORTER CONTACTED ANIMAS ALLEGING THAT THE DISPLAY SCREEN HAD BECOME DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349514 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1