FDA Adverse Event
Malfunction
Summary report: N
CONTOUR NEXT
MDR report key: 3872300
·
Received June 13, 2014
Report
- Report Number
- 1826988-2014-00200
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 1, 2014
- Report Date
- May 28, 2014
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K110894
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE ADVOCATE STATED THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 246MG/DL FROM THE CONTOUR NEXT LINK METER, AND ALSO RAN TESTS ON 3 OTHER METERS WITH READINGS OF 121, 122, 123, 124, 125 AND 127MG/DL. THE DIFFERENCES BETWEEN THE READINGS FALL IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCES CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE ADVOCATE WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349183 | CONTOUR NEXT | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7311 | 2LFEC05 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |