ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-16806
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Report Date
- June 9, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 08/07/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE BLACK BOX REVIEW SHOWS EAW-162 LOSS OF PRIME WARNING ASSOCIATED WITH A LOW NON-ZERO FORCE. AN EZPRIME SEQUENCE AND 24HR DURATION TEST WERE SUCCESSFULLY COMPLETED WITH NO LOSS OF PRIME WARNINGS BEING DUPLICATED. THE FORCE SENSOR CALIBRATION CHECK SHOWED THE PUMP IS DETECTING THE CORRECT FORCE. REMOVED PUMP COVER; NO CONTAMINATION OR DAMAGE TO THE FORCE SENSOR CIRCUIT WAS OBSERVED. UNIDENTIFIED FORCE LOW OBSERVED IN THE BLACK BOX, FORCE SENSOR CALIBRATION IN SPECIFICATIONS. THE BATTERY COMPARTMENT IS CRACKED BELOW BUMPER GRIP. RETURNED BATTERY CAP IS ABLE TO FULLY TIGHTEN TO THE PUMP AND WAS USED TO COMPLETE ALL TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349768 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |