RAD-8, HORIZONTAL
Report
- Report Number
- 2031172-2014-00052
- Event Type
- Death
- Date Received
- May 30, 2014
- Date of Event
- April 10, 2014
- Report Date
- April 30, 2014
- Manufacturer
- MASIMO CORPORATION
- Product Code
- DQA
- PMA / PMN Number
- K042536
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE RETURNED THAT WAS INVOLVED IN THE REPORTED EVENT WAS EVALUATED. A VISUAL INSPECTION OF THE EXTERNAL SURFACES OF THE DEVICE CONCLUDED NO EXTERNAL DAMAGE TO WAS OBSERVED. A SPO2 SIMULATOR WAS USED TO VERIFY THE FUNCTIONALITY OF SPO2 AND PULSE RATE, IT WAS FOUND TO BE FUNCTIONING AS DESIGNED. TESTING WAS ALSO PERFORMED USING A KNOWN GOOD SENSOR. TESTING SHOWED THE UNIT TO PROVIDE BOTH AUDIBLE AND VISUAL ALARM WHEN THE ALARM THRESHOLDS FOR LOW SPO2 WERE REACHED.
IT WAS REPORTED BY THE CUSTOMER THAT THIS UNIT DID NOT ALARM DURING A PT DE-SATURATION. THE CUSTOMER STATES THAT AFTER RECEIVING THE UNIT BACK FROM THE END USER, THEY TESTED THE UNIT WITH THEIR OWN CABLE AND SENSOR AND FOUND THAT THE UNIT WAS ALARMING CORRECTLY WHEN IT SURPASSED A SPO2 LIMIT THAT THEY PURPOSELY SET TO A LEVEL THAT ALLOWED THEM TO TEST IT. THE SENSOR USED DURING THE REPORTED EVENT IS NO LONGER AVAILABLE AS IT HAS BEEN DISCARDED BY THE END USER. THE LOT NUMBER OF THE SENSOR IS UNK AS WELL. ADD'L INFO REC'D FROM THE CUSTOMER ON (B)(6) 2014. THE CUSTOMER INDICATED THAT THE PT PASSED AWAY BECAUSE OF THIS INCIDENT; THE MOTOR THOUGH THE PT WAS ASLEEP UNTIL THE VENTILATOR STARTED TO ALARM; THE FATHER KNEW SOMETHING WAS WRONG; THE SENSOR WAS DISCARDED; THE SENSOR LOT NUMBER IS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319296 | RAD-8, HORIZONTAL | DQA | MASIMO CORPORATION | 22042 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 4 MO | Death | VENTILATOR |