FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3872003
·
Received June 13, 2014
Report
- Report Number
- 2531779-2014-16953
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Report Date
- June 10, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
Description of Event or Problem · 1
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 351084 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |