OXF ANATOMIC BRG RT LG SIZE 4
Report
- Report Number
- 0001825034-2014-05393
- Event Type
- Injury
- Date Received
- June 13, 2014
- Date of Event
- May 21, 2014
- Report Date
- March 4, 2016
- Manufacturer
- BIOMET UK LTD.
- Product Code
- NRA
- PMA / PMN Number
- PP010014
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
THIS FOLLOW-UP REPORT IS BEING FILED TO CORRECT INFORMATION.
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER WARNINGS, NUMBER 1 STATES, "IMPROPER SELECTION, PLACEMENT, POSITIONING, ALIGNMENT AND FIXATION OF THE IMPLANT COMPONENTS MAY RESULT IN UNUSUAL STRESS CONDITIONS WHICH MAY LEAD TO SUBSEQUENT REDUCTION IN THE SERVICE LIFE OF THE PROSTHETIC COMPONENTS."
EXAMINATION OF RETURNED DEVICE FOUND NO EVIDENCE OF PRODUCT NON-CONFORMANCE. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED DUE TO INSUFFICIENT INFORMATION.
IT WAS REPORTED THAT PATIENT UNDERWENT A RIGHT PARTIAL KNEE ARTHROPLASTY ON (B)(6) 2014. SUBSEQUENTLY, PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2014 DUE TO A POSTERIORLY DISLOCATED TIBIAL BEARING AND MIDFLEXION INSTABILITY. THE TIBIAL BEARING WAS REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350895 | OXF ANATOMIC BRG RT LG SIZE 4 | PROSTHESIS, KNEE | NRA | BIOMET UK LTD. | N/A | 2194070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |