FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3871814 · Received April 17, 2014

Report

Report Number
1720753-2014-03358
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 31, 2014
Report Date
April 17, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FILAMENT DRIVER PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A STATOR NOT ON ERROR MESSAGE. THIS ERROR WILL LIKELY CAUSE THE SYSTEM TO SHUT DOWN UN-COMMANDED, WHICH WOULD RESULT IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236201 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1