POWERPICC SOLO HF CATH W/SHER 3CG (TPS) STYLET 5F MAX
Report
- Report Number
- 3006260740-2014-00190
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 27, 2014
- Report Date
- March 27, 2014
- Manufacturer
- C.R. BARD, INC. (BASD)
- Product Code
- LJS
- PMA / PMN Number
- K091324
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NURSE
Narratives
THE COMPLAINT OF A LEAK IN THE CATHETER WAS CONFIRMED, AND THE CAUSE APPEARS TO BE USER RELATED. THE PRODUCT RETURNED FOR EVAL WAS A 5FR TRIPLE-LUMEN POWERPICC SOLO CATHETER. DISCOLORATION WAS SEEN IN THE TUBING, THROUGHOUT THE EXTENSION TUBING AND CATHETER SHAFT. THE CATHETER APPEARED TO BE TRIMMED AT THE 25CM DEPTH MARKING. A 0.2" LENGTH LONGITUDINAL SPLIT WAS OBSERVED BETWEEN THE 1CM AND 2 CM DEPTH MARKING. THE SAMPLE WAS PATENT TO INFUSION USING WATER FROM A 12CC SYRINGE WITH WATER LEAKING THROUGH THE SPLIT SITE WHEN THE GREY-LUERED LUMEN WAS FLUSHED. MICROSCOPIC EXAMINATION OF THE SPLIT SITE REVEALED A DULL AND GRANULAR FRACTURE SURFACE. THE OUTER SURFACE OF THE SPLIT SITE HAD A RIPPLE-LIKE SHAPE TO THE MATERIAL. THE CHARACTERISTICS OF THE OBSERVED DAMAGE ARE INDICATIVE OF A BURST IN THE LUMEN DUE TO OVER-PRESSURIZATION. THE IFU FOR THIS DEVICE PROVIDES ADEQUATE INSTRUCTION FOR INFUSING THIS DEVICE AS WELL AS FOR AVOIDING BURSTING THE LUMEN BY OVER-PRESSURIZATION. A LOT HISTORY REVIEW (LHR) OF REYB0126 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINT(S) FROM THIS LOT NUMBER.
PICC WAS FOUND PERFORATED AT 2CM MARK IN PT. LEAKING FROM THE INSERTION SITE. THE CATHETER WAS INSERTED TO THE 0CM MARK. PICC WAS REPLACED WITH A NEW DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236627 | POWERPICC SOLO HF CATH W/SHER 3CG (TPS) STYLET 5F MAX | LJS | C.R. BARD, INC. (BASD) | REYB0126 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |