FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3871600 · Received June 13, 2014

Report

Report Number
2531779-2014-16824
Event Type
Malfunction
Date Received
June 13, 2014
Report Date
June 9, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. IT WAS REPORTED THAT THE PUMP WAS LOSING POWER INTERMITTENTLY. IT WAS NOTED THAT THE POWER ISSUE OCCURRED DURING OR IMMEDIATELY AFTER A BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350540 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 20 YR