S-ROM M HEAD 36MM +3
Report
- Report Number
- 1818910-2014-20979
- Event Type
- Injury
- Date Received
- June 13, 2014
- Date of Event
- May 9, 2014
- Report Date
- February 5, 2015
- Manufacturer
- DEPUY IRELAND 9616671
- Product Code
- JDI
- PMA / PMN Number
- K120599
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- ATTORNEY
Narratives
(B)(4). THE INVESTIGATION HAS BEEN REOPENED DUE TO RECEIVING PART AND LOT NUMBER. DEPUY WILL NOTIFY THE FDA WHEN THE INVESTIGATION IS COMPLETE.
EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED OTHER REPORTS AGAINST THE FEMORAL HEAD AND/OR LINER. PER PROCEDURE, THIS DEVICE(S) IS EXEMPT FROM DEVICE HISTORY RECORD REVIEW. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED NO OTHER REPORTS AGAINST THE REMAINING PRODUCT/LOT CODE COMBINATION. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THE INVESTIGATION WILL BE REOPENED.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE AND/OR DHR REVIEW WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.
EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED OTHER REPORTS AGAINST THE FEMORAL HEAD AND/OR LINER. PER PROCEDURE, THIS DEVICE(S) IS EXEMPT FROM DEVICE HISTORY RECORD REVIEW. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED NO OTHER REPORTS AGAINST THE REMAINING PRODUCT/LOT CODE COMBINATION. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THE INVESTIGATION WILL BE REOPENED.
DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
LITIGATION ALLEGES THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION AND LARGE AMOUNTS OF TOXIC COBALT-CHROMIUM METAL IONS AND PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE AND BONE.
**UPDATE** 71/2016 - SALES REP REPORTED PATIENT WAS REVISED DUE TO PAIN AND PSEUDOTUMOR WITH MINIMAL METALLOSIS.
PFS RECEIVED. THE STEM IS BEING ADDED FOR THE ALLEGED HIGH METAL IONS (NO LABS RESULTS PROVIDED).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 349712 | S-ROM M HEAD 36MM +3 | HIP FEMORAL HEAD | JDI | DEPUY IRELAND 9616671 | 2653976 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | Required Intervention |