FDA Adverse Event Malfunction Summary report: N

SMALL BATTERY DRIVE II

MDR report key: 3871349 · Received June 13, 2014

Report

Report Number
2520274-2014-11687
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 29, 2014
Report Date
May 29, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS RETURNED FOR SERVICE HOWEVER DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE IMMERSION WHILE CLEANING (USER ERROR/MISUSE/ABUSE). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE SMALL BATTERY DRIVE DEVICE WAS MAKING AN UNUSUAL NOISE. THE EVENT WAS NOT RELATED TO SURGERY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350966 SMALL BATTERY DRIVE II INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS 1132

Patients

Seq Age Sex Outcome Treatment
1