FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3871344 · Received June 13, 2014

Report

Report Number
0001811755-2014-02138
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 19, 2014
Report Date
May 19, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED HEAT AND NOTED CORRODED DRIVESHAFT BEARINGS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350467 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1