LOANER SYSTEM 6 SAG SAW
Report
- Report Number
- 0001811755-2014-02139
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 27, 2014
- Report Date
- May 27, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED DISASSEMBLY OF THE TRIGGER WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH VISUAL INSPECTION. UPON DISASSEMBLY, IT WAS FOUND THAT THE TRIGGER HOUSING WAS CRACKED, WHICH CAN LEAD TO THE REPORTED EVENT AND CAN BE CAUSED BY A MATERIAL FATIGUE ISSUE OR IMPACT.
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
IT WAS REPORTED THAT UPON RECEIPT AT THE USER FACILITY, THE DEVICE DISASSEMBLED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT UPON RECEIPT AT THE USER FACILITY THE DEVICE DISASSEMBLED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350934 | LOANER SYSTEM 6 SAG SAW | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |