VASO VIEW HEMOPRO
Report
- Report Number
- 2242352-2014-00429
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 19, 2014
- Report Date
- March 19, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE BOOTS DISPLAYED EVIDENCE OF HEAT RELATED DAMAGE CONSISTENT WITH ACTIVATION OF THE DEVICE WHILE THE JAWS ARE IN DIRECT CONTACT. THERE WAS A CRACK ON THE TIP OF THE SILICONE BOOT ON THE COLD JAW. WHILE WE WERE UNABLE TO CONCLUSIVELY DETERMINE THE ROOT CAUSE, THIS PROBLEM IS CONSISTENT WITH AN IMPROPER INSERTION OF THE TOOL INTO THE CANNULA. BASED UPON THE EVAL RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION (WY-2013-05) ISSUED FOR THIS FAILURE MODE.(B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE SILICONE ON THE JAW OF THE VASO VIEW HEMOPRO WAS CRACKED. THE PA OBSERVED THIS WHEN THE DEVICE WAS IN THE PT'S LEG. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230646 | VASO VIEW HEMOPRO | ENDOSCOPIC VESSEL HARVESTING | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25092014 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |