HUT EXT DR FINAL ASSY-STANDARD
Report
- Report Number
- 1518293-2014-00043
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 17, 2014
- Report Date
- March 17, 2014
- Manufacturer
- LIEBEL-FLARSHEIM
- Product Code
- IXR
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
CUSTOMER CALLED TECH SUPPORT REPORTING THEY WERE NOT ABLE TO FLUORO OR TAKE RAD SHOTS AND WERE DECIDING IF THEY NEEDED TO HAVE A SERVICE ENGINEER VISIT THE SITE. CUSTOMER SAID THEY NEEDED TO SEE IF THEIR OWN BIOMED COULD FIX THE UNIT, AND WOULD CALL BACK IF NEEDED. ON A TECH SUPPORT FOLLOW UP CALL, CUSTOMER SAID THAT THE FSE (FIELD SERVICE ENGINEER) WAS NO LONGER ENDED FOR THIS UNIT. THEY DETERMINED THE PROBLEM WAS WITH THE MONITOR POWER SUPPLY AND HAD ORDERED A PART FROM LIEBEL FLARSHEIM TECH SUPPORT THEN CLOSED THE SERVICE TICKET. HOWEVER, THE CUSTOMER CALLED BACK (B)(4) SAYING THEY NOW WANTED THE FSE TO REPLACE THE POWER SUPPLY. FSE REPLACED THE POWER SUPPLY, BUT AFTER SAME TISSUE RE-OCCURRED, RE-INSTALLED THE CUSTOMERS ORIGINAL POWER SUPPLY. THE FSE FURTHER EVALUATED AND FOUND THE INTERMITTENT PROBLEM TO BE THE INFIRMED COMPUTER AND REPLACED THE CUSTOMERS COMPUTER WITH A LOANER. UNIT WAS CHECKED FOR PROPER OPERATION AND RETURNED TO THE CUSTOMER FOR SERVICE.
CUSTOMER REPORTS VIA PHONE THAT FLUORO FAILED DURING A PT PROCEDURE, STAFF BROUGHT A PORTABLE C-ARM FLUORO UNIT IN TO COMPLETE THE PROCEDURE. NO REPORTED INJURY. CUSTOMER PROVIDED NO FURTHER INFO ABOUT THE PT OR PROCEDURE, OTHER THEN TO SAY THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230641 | HUT EXT DR FINAL ASSY-STANDARD | IXR | LIEBEL-FLARSHEIM | HUT EXT DR FINAL A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |