HS III PROXIMAL SEAL SYTEM 3.8MM
Report
- Report Number
- 2242352-2014-00426
- Event Type
- Malfunction
- Date Received
- April 15, 2014
- Date of Event
- March 10, 2014
- Report Date
- March 19, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, SEAL AND ANCHOR WERE INSIDE THE LOADING DEVICE. THE BLUE SLIDE LOCK WAS LOCKED AND THE WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAIL TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. INTERNAL CORRECTIVE ACTION (WY-2013-35) ISSUED FOR THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY BYPASS PROCEDURE. THE HEARTSTRING III FAILED TO LOAD PROPERLY. THE SEAL REMAINED INSIDE THE LOADER. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230640 | HS III PROXIMAL SEAL SYTEM 3.8MM | ENDOSCOPIC VESSEL HARVESTING | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25092967 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |