FDA Adverse Event Malfunction Summary report: N

HIRES 90K

MDR report key: 3871177 · Received May 29, 2014

Report

Report Number
3871177
Event Type
Malfunction
Date Received
May 29, 2014
Date of Event
April 9, 2014
Report Date
May 29, 2014
Manufacturer
ADVANCED BIONICS CORPORATION
Product Code
MCM
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MI, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

A HIRES 90K ADVANTAGE COCHLEAR IMPLANT DEVICE WITH A HIFOCUS MID-SCALA ELECTRODE, WAS INITIALLY PLACED IN THE POCKET. THIS WAS LOADED ON THE INSERTION TOOL. AFTER THE INSERTION TOOL WAS DISCHARGED, IT WAS NOTICED THAT THE STYLET WAS NOT REMOVE BY THE INSERTION TOOL AS IT SHOULD. THE ELECTRODE WAS THEN WITHDRAWN AND INSPECTED UNDER THE MICROSCOPE, AND THERE WAS A POSSIBILITY OF A BREACH OF THE DISTAL TIP WHERE THE STYLET MAY HAVE PIERCED IT. THAT DEVICE WAS REMOVED FROM THE POCKET AND REMOVED AND RETURNED TO THE COMPANY, AND WE ELECTED TO PROCEED WITH THE BACKUP DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
317305 HIRES 90K IMPLANT, COCHLEAR MCM ADVANCED BIONICS CORPORATION CI-1500-04 *

Patients

Seq Age Sex Outcome Treatment
1 59 YR