FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3870896 · Received June 12, 2014

Report

Report Number
3004209178-2014-85600
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
February 17, 2014
Report Date
May 29, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K053177
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING THE REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND, AND SOME MOTOR ERROR ALARMS WERE REGISTERED IN THE ALARM HISTORY. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347174 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-712EWS

Patients

Seq Age Sex Outcome Treatment
1