ETS LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-03961
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Report Date
- April 30, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: DAMAGED YOKE TEETH. THE ANALYSIS RESULTS FOUND THAT AN ATS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD LOADED IN THE DEVICE. THE CLAMPING MECHANISM WAS NOTED TO BE DAMAGED. NO FUNCTIONAL TEST WAS PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE YOKE TEETH WERE FOUND BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE TEETH BECAME BROKEN, THIS DAMAGE CAN OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE CLOSING TRIGGER WHEN THE JAWS ARE CLAMPED ON TISSUE THICKER THAN INDICATED. IN ADDITION, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).
THE DEVICES RECEIVED WITH NO FURTHER INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347136 | ETS LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |