FDA Adverse Event Malfunction Summary report: N

ETS LINEAR CUTTER

MDR report key: 3870857 · Received June 12, 2014

Report

Report Number
3005075853-2014-03961
Event Type
Malfunction
Date Received
June 12, 2014
Report Date
April 30, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: DAMAGED YOKE TEETH. THE ANALYSIS RESULTS FOUND THAT AN ATS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD LOADED IN THE DEVICE. THE CLAMPING MECHANISM WAS NOTED TO BE DAMAGED. NO FUNCTIONAL TEST WAS PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE YOKE TEETH WERE FOUND BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE TEETH BECAME BROKEN, THIS DAMAGE CAN OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE CLOSING TRIGGER WHEN THE JAWS ARE CLAMPED ON TISSUE THICKER THAN INDICATED. IN ADDITION, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).

Description of Event or Problem · 1

THE DEVICES RECEIVED WITH NO FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347136 ETS LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK