FDA Adverse Event Malfunction Summary report: N

NA

MDR report key: 3870787 · Received June 12, 2014

Report

Report Number
0001056128-2014-00072
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
May 15, 2014
Report Date
May 19, 2014
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
Product Code
HRX
PMA / PMN Number
K012635
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT DEVICE WAS RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS (SSS) FOR AN EVALUATION. VISUAL INSPECTION REVEALED BIOLOGICAL MATERIAL ON THE DISTAL TIP OF THE DEVICE, INDICATING THE DEVICE WAS USED CLINICALLY. THE DISTAL TIP OF THE INNER SHAFT WAS OBSERVED TO BE BROKEN AND DETACHED. THE DISTAL TIP OF THE INNER SHAFT WAS NOT RETURNED FOR EVALUATION. GALLING WAS OBSERVED ON THE INSIDE THE DISTAL TIP OF THE OUTER SHAFT. THE RESULTS OF THE INVESTIGATION PERFORMED INDICATED THAT THE RETURNED DEVICE WAS OBSERVED TO HAVE A BROKEN/DETACHED DISTAL TIP ON THE INNER SHAFT, SUBSTANTIATING THE FACILITY¿S REPORTED ISSUE. BASED ON THE DAMAGE ON THE DISTAL TIPS, AND THE GALLING INSIDE THE DISTAL TIP OF THE OUTER SHAFT, THE MOST LIKELY CAUSES FOR THE REPORTED ISSUE WERE DETERMINED TO BE: 1) THE ARTHROSCOPIC SHAVER MAY HAVE COME INTO CONTACT WITH STAPLES, CLIPS OR ANOTHER METAL OBJECT, RESULTING IN DAMAGE TO THE BLADE; AND/OR 2) EXCESSIVE LATERAL FORCES EXERTED ON THE DEVICE DURING CLINICAL USE MAY HAVE CAUSED THE TEETH OF THE INNER SHAFT TO CONTACT THE CUTTING EDGES OF THE OUTER SHAFT, CAUSING THE TEETH TO CATCH ON THE OUTER SHAFT AS THE DEVICE ROTATED, AND THE OBSERVED GALLING AND DAMAGE TO THE DEVICE. THE INSTRUCTIONS FOR USE (IFU) FOR ARTHROSCOPIC SHAVERS (AS) STATES THE FOLLOWING: "DO NOT ALLOW THE ARTHROSCOPIC SHAVER TO COME INTO CONTACT WITH STAPLES, CLIPS OR ANY METAL OBJECT TO AVOID DAMAGE TO THE BLADE AND POSSIBLE PATIENT INJURY." "CAREFUL HANDLING OF THE INSTRUMENT IS NECESSARY TO AVOID DAMAGE OR BREAKAGE AS A RESULT OF EXCESSIVE FORCE." "DO NOT APPLY EXCESSIVE PRESSURE OR ¿SIDE-LOAD¿ THE BLADE DURING USE. SIDE-LOADING DOES NOT IMPROVE THE PERFORMANCE OF THE INSTRUMENT, CAN DULL THE BLADE, AND/OR PRODUCE METAL PARTICULATES." PRODUCT SURVEILLANCE WILL CONTINUE TO MONITOR COMPLAINTS OF THIS TYPE FOR ADVERSE TRENDS. THIS REPORT IS BEING FILED AS A MALFUNCTION DUE TO SSS BEING IN A 2 YEAR REPORTING CYCLE DUE TO MDR 0001056128-2014-00007 FILED ON (B)(4) 2014 WHERE TIP BROKE OFF AND PROCEDURE WENT FROM ARTHROSCOPIC TO OPEN IN ORDER TO REMOVE THE TIP. THE REPORTED EVENT IS NOT OCCURRING MORE FREQUENTLY OR WITH GREATER SEVERITY THAN IS USUAL FOR THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT, "DURING SURGERY, A PIECE OF THE SHAVER BROKE OFF AND WAS RETRIEVED." THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347625 NA ARTHROSCOPE HRX STRYKER SUSTAINABILITY SOLUTIONS LAKELAND 375-545-000 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1