FDA Adverse Event Malfunction Summary report: N

3.0MM TI CANNULATED SCREW INSTRUMENT & IMPLANT SET

MDR report key: 3870709 · Received June 12, 2014

Report

Report Number
2520274-2014-11799
Event Type
Malfunction
Date Received
June 12, 2014
Report Date
October 29, 2012
Manufacturer
SYNTHES USA
Product Code
FSM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. PRODUCT CODE NOT REQUIRED. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED / EXPLANTED. DEVICE WAS RECEIVED FOR EVALUATION (DATE UNSPECIFIED). IF 145.17 IS A SET CONFIGURATION, THEN N/A WOULD BE AN APPROPRIATE CLASSIFICATION AS SETS THEMSELVES ARE NOT REGULATED, ONLY THE COMPONENTS CONTAINED THEREIN. INVESTIGATION IS ON-GOING. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. NO CONCLUSION CAN BE DRAWN. REVIEW OF MANUFACTURING RECORDS HAS BEEN REQUESTED. RESIDUAL TESTING WAS PERFORMED ON THE SET. THE RESULTS SHOWED ON (B)(6) 2014 THAT THE IMPLANTS ARE CYTOTOXIC AND FURTHERMORE SOME NON-VOLATILE MATERIALS (YET TO BE IDENTIFIED) ALSO ARE BEING IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 3.0 MM TI CANNULATED SCREW INSTRUMENT AND IMPLANT SET (P/N 145.17) WAS SUBMERGED IN WATER AFTER A STORM, HURRICANE SANDY. THE SALES CONSULTANT¿S CAR AND THE SET WERE COMPLETELY UNDER WATER FOR A FEW HOURS, POSSIBLY LONGER; SUBSEQUENTLY, THE SET WAS SOAKED. SALES CONSULTANT ALSO MENTIONED THAT WHEN HE LOOKED INSIDE THE SET THERE WAS SAND/SEDIMENT/MUD AND REMEMBERS SOME OXIDATION/RUST ON PARTS OF THE SET (HE COULDN¿T BE SPECIFIC). THE SET WAS SUPPOSED TO BE SET ASIDE TO BE DISPOSED OF BUT WAS INADVERTENTLY MISSED. WHEN RETURNED TO THE SYNTHES FACILITY, THE SET WAS DECONTAMINATED AND SENT OUT TO BE USED BY A HOSPITAL. REPORTEDLY, THE FIRST TIME IT WAS SHIPPED OUT, THE SET WAS NOT USED AND WAS SENT BACK UNUSED. THE SET WAS THEN DECONTAMINATED AGAIN WHEN IT WAS RETURNED AND EVENTUALLY SENT OUT TO A HOSPITAL WHERE IT WAS USED. THE SET WAS RETURNED AND USED 4 TIMES TOTAL AFTER BEING INVOLVED IN THE HURRICANE. THIS REPORT IS 1 OF 1 FOR COM-(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348341 3.0MM TI CANNULATED SCREW INSTRUMENT & IMPLANT SET TRAY, SURGICAL INSTRUMENT FSM SYNTHES USA

Patients

Seq Age Sex Outcome Treatment
1