PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 3870690
·
Received June 12, 2014
Report
- Report Number
- 2955842-2014-03618
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 26, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS BROKEN. THE PITCH DOWN CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, IT WAS NOTED THAT A BROKEN CABLE WAS OBSERVED ON THE PK DISSECTING FORCEPS INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.