V60 VENTILATOR
MDR report key: 3870568
·
Received June 12, 2014
Report
- Report Number
- 2031642-2014-00486
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Report Date
- June 11, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
FACTORY ANALYSIS OF THE MOTOR CONTROLLER AND POWER MANAGEMENT PCB BOARDS REVEALED SHORTED COMPONENTS WHICH MAY RESULT IN A LOSS OF THE 12 VOLT SUPPLY. IF THE NOTED FAILURE WERE TO OCCUR DURING OPERATION OF THE DEVICE DURING USE IN NORMAL VENTILATION, A VENT INOPERATIVE CONDITION COULD OCCUR. A VENT INOP CONDITION DURING OPERATION WILL RESULT IN A VISUAL AND AUDIBLE ALARM AND PRECLUDES CONTINUED SAFE OPERATION OF THE VENTILATOR. IF IN USE, THE PATIENT MUST BE PLACED ON ANOTHER MEANS OF VENTILATORY SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348012 | V60 VENTILATOR | VENTILATOR, CONTINUOUS | MNT | RESPIRONICS CALIFORNIA, INC. | V60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|