COREVALVE
Report
- Report Number
- 2025587-2014-00377
- Event Type
- Injury
- Date Received
- June 12, 2014
- Date of Event
- May 1, 2014
- Report Date
- March 5, 2019
- Manufacturer
- MEDTRONIC HEART VALVES DIVISION
- Product Code
- NPT
- PMA / PMN Number
- P130021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT ANALYSIS: THE PRODUCT REMAINS IMPLANTED, THEREFORE NO PRODUCT ANALYSIS CAN BE PERFORMED. CONCLUSION: BASED ON THE INFORMATION PROVIDED, THE REPORTED POSITIONING ISSUE THAT REQUIRED REINTERVENTION COULD POSSIBLY BE RELATED TO PATIENT ANATOMY AND/OR IMPLANT POSITION. POSITIONING IS DEPENDENT ON PATIENT ANATOMY AS WELL AS USER TECHNIQUE. POTENTIAL FACTORS THAT CAN INFLUENCE IMPLANT POSITION INCLUDE TENSION APPLIED ON THE DELIVERY CATHETER SYSTEM (DCS) DURING POSITIONING, CALCIFICATION LEVELS IN THE NATIVE VESSEL AND COMPLIANCE OF THE AORTA AND NATIVE VESSELS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO ADVERSE PATIENT EFFECTS REPORTED. THE ROOT CAUSE FOR THIS REPORT COULD NOT BE ESTABLISHED FROM THE INFORMATION AVAILABLE. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS TRANSCATHETER BIOPROSTHETIC VALVE, THE FIRST VALVE WAS NOT FULLY DEPL OYED AND IT WAS FELT THAT THE ANGLE OF IMPLANTATION WAS SUBOPTIMAL IN THE LEFT CORONARY REGION. THE VALVE COULD NOT BE RETRIEVED FROM THE PATIENT AND WAS DEPLOYED IN THE AORTIC ARCH. A SECOND VALVE WAS IMPLANTED SUCCESSFULLY WITH NO ADVERSE PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348011 | COREVALVE | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | MEDTRONIC HEART VALVES DIVISION | MCS-P3-3143 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00085 YR | Required Intervention |