SEE H-10
Report
- Report Number
- 2023826-2014-00438
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P880091
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
BRAND NAME: SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIC®) WITH TORIC. (B)(4). EVALUATION, RESULTS: VISUAL INSPECTION OF THE PRODUCT SHOWED THE LENS WAS CUT INTO THREE PIECES AND HALF OF THE OPTIC AND A PIECE OF THE OPTIC WAS TORN OFF AND MISSING. CONCLUSION: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED, WHICH WAS SUBSEQUENTLY CLOSED. THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE CUSTOMER REPORTED THAT THE AA4203TL SILICONE SINGLE PIECE LENS WITH TORIC, TORE AS THE SURGEON PLACED IT IN THE EYE. THE LENS WAS REMOVED WITHOUT ALTERING THE INICISION AND ANOTHER TORIC LENS WAS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348492 | SEE H-10 | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | AA4203TL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | MSI-TR INJECTOR MODEL. LOT NUMBER UNKNOWN| MTC-60CFP CARTRIDGE, LOT NUMBER UNKNOWN |