FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3870466 · Received June 12, 2014

Report

Report Number
2531779-2014-16769
Event Type
Malfunction
Date Received
June 12, 2014
Report Date
June 4, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE TORN AT THE UP ARROW BUTTON. ALL OF THE KEYPAD BUTTONS RESPONDED APPROPRIATELY. THE PUMP WAS OPENED AND CONTAMINATION WAS FOUND UNDER THE UP ARROW AND CONTRAST BUTTON CONTACTS. EVALUATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED ON THE THREADS AND BATTERY CAP WAS DAMAGED STRIPPED. EVALUATION ALSO REVEALED A DIM, FADING AND DISCOLORED DISPLAY SCREEN. UNRELATED TO THESE ISSUES, THE AUDIO BOLUS BUTTON COVER WAS TORN BUT THE BUTTON RESPONDED NORMALLY. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED EVIDENCE OF CONTAMINATION UNDER THE UP ARROW AND CONTRAST KEYPAD BUTTON CONTACTS. EVALUATION ALSO REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED ON THE THREADS AND THE BATTERY CAP WAS STRIPPED. IN ADDITION, THE DISPLAY SCREEN WAS DIM, FADING AND DISCOLORED. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348601 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 22 YR