FDA Adverse Event Injury Summary report: N

SCREW M6*18 F/03.401.072

MDR report key: 3870363 · Received June 12, 2014

Report

Report Number
2520274-2014-11754
Event Type
Injury
Date Received
June 12, 2014
Date of Event
May 15, 2014
Report Date
May 15, 2014
Manufacturer
SYNTHES (USA)
Product Code
HWC
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DATE OF BIRTH REPORTED ONLY AS 1954. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

A MANUFACTURING INVESTIGATION WAS CONDUCTED. THE REPORT INDICATES THAT: COMPLAINED ARTICLE FORWARDED TO THE LEGAL MANUFACTURER FOR EVALUATION. THE INVESTIGATION HAS SHOWN THAT SCREW HEAD IS BROKEN OFF FROM THE SCREW. THE ARTICLES WHERE ANALYZED FOR CONFORMANCE TO PRINT SPECIFICATIONS AS WELL AS THE DEVICE HISTORY RECORD WAS RESEARCHED. NO ABNORMAL FINDINGS WERE IDENTIFIED. UNFORTUNATELY WE ARE NOT ABLE TO DETERMINE THE EXACT CAUSE WHICH HAS LED TO THIS OCCURRENCE. WE CAN ONLY ASSUME THAT A MECHANICAL OVERLOAD SITUATION HAS LED TO THE BREAKAGE. NO PRODUCT FAILURE COULD BE DETECTED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED A PATIENT WITH PRIMARY TOTAL SHOULDER PROSTHESIS RETURNED WITH PAIN. AN ARTHROSCOPY WAS PERFORMED (B)(6) 2014 WHICH SHOWED SIGNS OF INFECTION AND A LOOSE GLENOID. ANTIBIOTICS WERE PRESCRIBED AND A REOPERATION WAS PLANNED TO EXTRACT EVERYTHING AND CHANGE TO AN ARTHREX PROSTHESIS, WHICH THEY NORMALLY USE. THE EPOCA STEM (UN-CEMENTED) WAS IMPOSSIBLE TO EXTRACT, THE INSTRUMENT BROKE AND THEY HAD TO CHANGE THE PLAN. THE DECISION WAS TO PUT IN A NEW EPOCA GLENOID, ECCENTER AND HEAD. THIS RESULTED IN A 1 HOUR DELAY, THE PATIENT SLEPT IN THE OPERATING ROOM WHILE THE SURGEON WAITED FOR THE EQUIPMENT. A NEW GLENOID,,ALL POLY,WAS IMPLANTED TOGETHER WITH A REVISION ECCENTER AND HEAD. THE REASON FOR THE FAILURE WAS THE SCREW (03.401.077) IN THE EXTRACTION INSTRUMENT (03.401.072) THAT BROKE. THIS REPORT IS FOR THE SCREW IN THE INSTRUMENT THAT BROKE. THIS IS REPORT 1 OF 5 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347114 SCREW M6*18 F/03.401.072 SCREW, FIXATION, BONE HWC SYNTHES (USA)

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention