ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-16761
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Report Date
- June 9, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/08/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THERE WAS EVIDENCE OF MOISTURE BEHIND THE DISPLAY LENS. THE LEAK TEST FAILED DUE TO A DISPLAY LENS LEAK. MOISTURE WAS OBSERVED ON THE PUMP'S INTERNAL COMPONENTS. UNRELATED TO THE INITIAL COMPLAINT, THE PUMP POWERED ON TO A BLANK DISPLAY WITH FUNCTIONING AUDITORY AND VIBRATORY SETTINGS. THE LEAK TEST FAILED DUE TO AN AUDIO BOLUS BUTTON AND KEYPAD LEAK. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY W/ MOISTURE) ISSUE. IT WAS REPORTED THAT THERE WAS MOISTURE BEHIND THE DISPLAY SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 348298 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |