FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3870053 · Received June 12, 2014

Report

Report Number
1416980-2014-18762
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
May 18, 2014
Report Date
May 18, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, THE HOME PATIENT (HP) REPORTED CONNECTING TO THE HOMECHOICE CASSETTE WITHOUT THE LINE BEING PROPERLY PRIMED. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

WHILE TROUBLESHOOTING FOR A SYSTEM ERROR 2240 (AIR IN LINE) ALARM ON A HOMECHOICE DEVICE, A TECHNICAL SERVICE REPRESENTATIVE (TSR) DISCOVERED THAT A HOME PATIENT CONNECTED TO THE HOMECHOICE CASSETTE DURING THE PRIMING PHASE OF AUTOMATED PERITONEAL DIALYSIS (APD) THERAPY. THE TSR EXPLAINED PROPER PROCEDURES WITH THE PATIENT AND INSTRUCTED THE PATIENT TO RESTART THERAPY USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348258 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 84 YR HOMECHOICE