MONOPOLAR CURVED SCISSORSINSTRUMENT
Report
- Report Number
- 2955842-2014-03593
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 15, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE SYSTEM FOR LATEX CUT TEST. THE SCISSORS DID NOT CUT CLEANLY THROUGH .006 OF LATEX. THE LATEX SNAGGED AT THE SCISSOR TIPS. THE BLADE EDGES WERE NOT DAMAGED BUT THE EDGES EXHIBITED WEAR AT THE TIPS, NEGATIVELY AFFECTING CUT PERFORMANCE. AN ADDITIONAL OBSERVATION NOT REPORTED WAS TUBE EXTENSION HAD PAD PRINTING REMOVED. THE TUBE EXTENSION EXHIBITED PAD PRINTING REMOVED MEASURING .042 LOCATED AT THE TUBE EXTENSION AND MAIN TUBE INTERFACE. FAILURE ANALYSIS CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING / MISUSE. AN ADDITIONAL OBSERVATION NOT REPORTED WAS THE BANANA PLUG WAS BENT TO THE SIDE. THERE WAS NO DAMAGE TO THE BACK END. FAILURE ANALYSIS CONCLUDED THE DAMAGE WAS LIKELY DUE TO MISHANDLING / MISUSE. NO OTHER DAMAGE WAS FOUND.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS DULL. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347731 | MONOPOLAR CURVED SCISSORSINSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-12 | M10140123 218 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |