FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3870007
·
Received April 28, 2014
Report
- Report Number
- 1720753-2014-03713
- Event Type
- Malfunction
- Date Received
- April 28, 2014
- Date of Event
- April 4, 2014
- Report Date
- April 28, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HV (HIGH VOLTAGE) CABLE AND IGBT PCB ASSEMBLY WERE EVALUATED AND IDENTIFIED AS REQUIRING REPLACEMENT. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFORMATION IS UNAVAILABLE AT THIS TIME AND NO ADDITIONAL SERVICE INFORMATION WAS PROVIDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A 'KV ON IN ERROR' MESSAGE DISPLAYED. THIS ERROR IS LIKELY TO RESULT IN AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254275 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |