GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HV (HIGH VOLTAGE) CABLE AND IGBT PCB ASSEMBLY WERE EVALUATED AND IDENTIFIED AS REQUIRING REPLACEMENT. NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFORMATION IS UNAVAILABLE AT THIS TIME AND NO ADDITIONAL SERVICE INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A 'KV ON IN ERROR' MESSAGE DISPLAYED. THIS ERROR IS LIKELY TO RESULT IN AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254275 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800