FDA Adverse Event Malfunction Summary report: N

COBAS E411 SYSTEM

MDR report key: 3869460 · Received June 12, 2014

Report

Report Number
1823260-2014-04299
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
May 22, 2014
Report Date
July 15, 2014
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JLB
PMA / PMN Number
K961481
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION FOR FURTHER INVESTIGATION WAS REQUESTED BUT WAS NOT PROVIDED. BASED ON THE PROVIDED CALIBRATION DATA, GENERAL REAGENT ISSUES WERE NOT SUSPECTED.

Additional Manufacturer Narrative · 1

INFORMATION CONCERNING THE SPECIFIC PART NUMBER INVOLVED IN THIS EVENT WAS REQUESTED, BUT WAS NOT PROVIDED. THIS EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

THE CUSTOMER RECEIVED QUESTIONABLE CK-MB - THE MB ISOENZYME OF CREATINE KINASE STAT AND TROPONIN I STAT RESULTS FOR ONE PATIENT. THE INITIAL CKMB RESULT WAS 300 NG/ML WITH A DATA FLAG. THE INITIAL TROPONIN I RESULT WAS 25.00 NG/ML WITH A DATA FLAG. THE REPEAT CKMB RESULT WHEN DILUTED 1:10 WAS 41.28 NG/ML WITH A DATA FLAG. THE REPEAT TROPONIN I RESULT WHEN DILUTED 1:10 WAS 23.21 NG/ML WITH A DATA FLAG. THESE RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE SECOND REPEAT CKMB RESULT WHEN DILUTED 1:10 WAS 404.7 NG/ML WITH A DATA FLAG. THE SECOND REPEAT TROPONIN I RESULT WHEN DILUTED 1:10 WAS 197.9 NG/ML WITH A DATA FLAG. THESE RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE CKMB REAGENT LOT NUMBER WAS 174105 AND THE TROPONIN I REAGENT LOT NUMBER WAS 174423. THE EXPIRATION DATES WERE REQUESTED, BUT WERE NOT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348284 COBAS E411 SYSTEM IMMUNOCHEMISTRY ANALYZER JLB ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1 069 YR PLAVIX| CLEXANE| ALDACTONE| RIVOTRIL